Saikang Ventilator






Saikang Ventilator
Description
Ventilator
Colour LCD display, touch screen setting, big font display, free changeable viewing angles
Various ventilation modes SIPPV, SIMV, PCV, SPONT, MANUAL, CPAP
Ventilation parameters PEEP, SIGH, IRV, Inspiratory platform
Tidal volume 50-1500 ml
Rates 1?99bpm
SIM Vrate 2?20bpm
I:E 4:1?1:8
Trigger sensitivity?-1.0?1.0kPa
PEEP 0?25cmH2O
Pressure control 3?50cmH2O
SIGNS 1.5 times inspiration time (60-120 times adjustable)
Inspiration platform 0?50%
Oxygen concentration 48%?100%
Pressure limit 5?60cmH2O
Parameters for monitoring tidal ventilation volume, ventilation volume, IPPV rate, total respiratory rate, peak airway pressure, pressure - time waveform, flow rate - time waveform, inspiratory pressure trigger, Inspiratory platform, Oxygen concentration, battery capacity
Security alarm system
Upper airway pressure alarm upper limit setting range 0.3?6.0 kPa Lower limit setting range 0.2?5.0 kPa
Ventilation volume alarm per minute Upper limit setting range 3.0?30L/min Lower limit setting range 1.0?10L/min
Adjustment range of the upper limit of the Vt alarm 10?2000ml Adjustment range of the lower limit 0?1800 ml
Sustained high pressure alarm: it gives an alert when the stress has been consistently above 2.5 kPa.
Smothering alarm it gives an audible and luminous alarm if the tidal volume has not been entered for 15 seconds
Power failure alarm
Power?AC 220V±10% 50Hz±1Hz
UPS, Storage battery
Battery undervoltage alarm
Power loss alarm
Intubation out of alert status






How do you want to proceed?
How do you want to proceed?
Description
Ventilator
Colour LCD display, touch screen setting, big font display, free changeable viewing angles
Various ventilation modes SIPPV, SIMV, PCV, SPONT, MANUAL, CPAP
Ventilation parameters PEEP, SIGH, IRV, Inspiratory platform
Tidal volume 50-1500 ml
Rates 1?99bpm
SIM Vrate 2?20bpm
I:E 4:1?1:8
Trigger sensitivity?-1.0?1.0kPa
PEEP 0?25cmH2O
Pressure control 3?50cmH2O
SIGNS 1.5 times inspiration time (60-120 times adjustable)
Inspiration platform 0?50%
Oxygen concentration 48%?100%
Pressure limit 5?60cmH2O
Parameters for monitoring tidal ventilation volume, ventilation volume, IPPV rate, total respiratory rate, peak airway pressure, pressure - time waveform, flow rate - time waveform, inspiratory pressure trigger, Inspiratory platform, Oxygen concentration, battery capacity
Security alarm system
Upper airway pressure alarm upper limit setting range 0.3?6.0 kPa Lower limit setting range 0.2?5.0 kPa
Ventilation volume alarm per minute Upper limit setting range 3.0?30L/min Lower limit setting range 1.0?10L/min
Adjustment range of the upper limit of the Vt alarm 10?2000ml Adjustment range of the lower limit 0?1800 ml
Sustained high pressure alarm: it gives an alert when the stress has been consistently above 2.5 kPa.
Smothering alarm it gives an audible and luminous alarm if the tidal volume has not been entered for 15 seconds
Power failure alarm
Power?AC 220V±10% 50Hz±1Hz
UPS, Storage battery
Battery undervoltage alarm
Power loss alarm
Intubation out of alert status
Questions and answers